FDA RECALL RECORD · Z-1860-2025

DORO Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002) Recall

DORO¿ Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002)

FDA recall Class II

Ongoing

FDA lists this Class II recall for DORO Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1860-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 14, 2025
FDA report date
Jun 4, 2025
Quantity in commerce
217 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

DORO Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002)

Affected model & UDI checker

Search the exact identifiers published for Z-1860-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedDORO Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002)Check source notice04250435506738
FDA code information

Model/Item No: 1204.002; UDI-DI: 04250435506738; Serial numbers: 1102010, 1201002, 1201004, 1201005, 1201006, 1201008, 1201009, 1201011, 1206012, 1206013, 1206014, 1206015, 1206016, 1206018, 1310001, 1310002, 1310003, 1312001, 1312002, 1312003, 1312004, 1312005, 1312006, 1312007, 1312008, 1312009, 1312010, 1407011, 1407012, 1407013, 1407014, 1407015, 1407016, 1407017, 1407018, 1407019, 1407020, 1501026, 1501027, 1501028, 1501029, 1501030, 1503031, 1503032, 1503033, 1503034, 1503035, 1503035, 1503036, 1503037, 1503038, 1503039, 1503040, 1505041, 1505042, 1505043, 1505044, 1505045, 1505046, 1505047, 1505048, 1505049, 1505050, 1506051, 1506052, 1506053, 1506054, 1507055, 1509056, 1509057, 1509058, 1509059, 1509060, 1509061, 1509062, 1509063, 1509064, 1509065, 1605066, 1605067, 1605068, 1605069, 1605070, 1605071, 1605072, 1605073, 1605074, 1605075, 1610078, 1610079, 1610080, 1610081, 1610082, 1610083, 1610084, 1610085, 1610086, 1708001, 1708002, 1708003, 1708004, 1708005, 1708006, 1708007, 1708008, 1708009, 1708010, 1801011, 1801012, 1801013, 1802014, 1802015, 1802016, 1802017, 1802018, 1802019, 1802020, 1805021, 1805022, 1805023, 1805024, 1805025, 1806026, 1806027, 1806028, 1806029, 1806030, 1807031, 1807032, 1807033, 1807034, 1811036, 1811037, 1811038, 1811039, 1811040, 1902001, 1902002, 1902003, 1902004, 1902005, 1902006, 1902007, 1902008, 1902009, 1902010, 1910001, 1910002, 1910003, 1910004, 1910005, 1910006, 1910007, 1910008, 1910009, 1910010, 2007001, 2007002, 2007003, 2007004, 2007005, 2007006, 2007007, 2007008, 2007009, 2007010, 2010001, 2010002, 2010003, 2010004, 2012001, 2012002, 2012004, 2012006, 2012007, 2012008, 2012009, 2012010, 2102001, 2104001, 2104002, 2104003, 2104004, 2104005, 2107001, 2107002, 2107006, 2107007, 2112001, 2112002, 2112003, 2202001, 2202002, 2202003, 2202004, 2202005, 2205001, 2205002, 2212001, 2212002, 2212003, 2212004, 2212005, 2301001, 2401001, 2401002, 2401003, 2401004, 2401005, 2401006, 2401007, 2405001, 2405002, 2405003, 2405004, 2408001, 2408002, 2408003, 2409001, 2409002, 2409003, 2409004;

View the original FDA recall notice

Why was this product recalled?

Potential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of the navigation system cannot be excluded.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Pro-Med Instruments Gmbh take?

On March 14, 2025 URGENT FIELD SAFETY NOTICE letters were emailed to distributors and customers. 5. Actions to be taken by the Customer/User: If you receive this Field Safety Notice from pro med instruments GmbH, part of Black Forest Medical Group, you have been identified as a hospital or user that have been supplied with the concerned product. The hospitals/ users who are in possession with the concerned product are kindly requested to - review this notification and ensure that all your affected products are identified and removed from use. - discontinue the use of the concerned product. - complete the attached Acknowledgement and Receipt Form , select your desired corrective action type (repair by the manufacturer or repair on-site) and return it by fax or email to pro med instruments GmbH, part of Black Forest Medical Group to confirm receipt by March 28th, 2025, at the latest (see contact information under section 9. below). Repairs will be available from the 4th of April 2025 and can be conducted after receipt of the completed Acknowledgement and Receipt Form . If, after reviewing this notification, you have any further questions or queries please consult your DORO sales representative. 8. Type of Action by pro med instruments, part of Black Forest Medical Group: Immediate actions: " Identification of customers/ hospitals/ users with concerned products " Recall of concerned products " Information to relevant national competent authorities Corrective Actions: " Ensure that the affected devices in the field have been separated " Restore product compatibility of separated devices within the field through free-of-charge upgrades of recalled products by: o upgrading of returned devices by the manufacturer's service department o provision of spare parts for upgrading the returned devices by the user or an authorized DORO sales representative on-site " Ensure that there are no more defective products and associated subcomponents in the field.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology