Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1861-2025

Plus Initial Drill, Article Number: 103.170; Recall

Plus Initial Drill, Article Number: 103.170;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Plus Initial Drill, Article Number: 103.170;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1861-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 14, 2025
FDA report date
Jun 4, 2025
Quantity in commerce
288 units ($& US, 241 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Plus Initial Drill, Article Number: 103.170;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Article Number: 103.170; UDI-DI: (01)7898237561967(11)250204(17)400204(10)LNWT8; Lot Number: LNWT8

View the original FDA recall notice

Why was this product recalled?

The referred batch was produced without the laser engraving of 7 mm.

What actions did Straumann USA LLC take?

On April 14, 2025 customers were notified via phone with a follow-up letter sent on April 28, 2024. Both contacts included the following: Action to be taken: 1. If article 103.170 from lot LNWT8 is still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 103.170 from Lot LNWT8 is still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. transmission of the Field Safety Notice This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice