IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set. Recall
IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
FDA recall Class I
Ongoing
FDA lists this Class I recall for IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1869-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2025
- FDA report date
- Jun 18, 2025
- Quantity in commerce
- 14,280 units (724 cases)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
Affected model & UDI checker
Search the exact identifiers published for Z-1869-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| SET-0014-20 | IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set. | Check source notice | Not specified |
| Not specified | IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set. | Check source notice | 20811505030034 |
FDA code information
Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015.
Why was this product recalled?
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
What actions did Fresenius Kabi USA, LLC take?
Consignees received an "URGENT - VOLUNTARY RECALL" notification via email dated 5/11/25. This notification instructs consignees to discontinue use and distribution of affected devices immediately. Consignees are to check their inventory for affected devices and place them in quarantine until product can be returned for replacement. The provided recall notification is to be distributed to users of the affected devices. Consignees are instructed to return the completed Customer Reply Form via fax or email to facilitate return of affected devices for replacement. Consignees with any questions can contact Fresenius Kabi by phone at 855-354-6387x1 or by email at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1869-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026