FDA lists this Class I recall for IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1869-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 12, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
14,280 units (724 cases)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.

Affected model & UDI checker

Search the exact identifiers published for Z-1869-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SET-0014-20IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.Check source noticeNot specified
Not specifiedIVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.Check source notice20811505030034
FDA code information

Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015.

View the original FDA recall notice

Why was this product recalled?

Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.

What actions did Fresenius Kabi USA, LLC take?

Consignees received an "URGENT - VOLUNTARY RECALL" notification via email dated 5/11/25. This notification instructs consignees to discontinue use and distribution of affected devices immediately. Consignees are to check their inventory for affected devices and place them in quarantine until product can be returned for replacement. The provided recall notification is to be distributed to users of the affected devices. Consignees are instructed to return the completed Customer Reply Form via fax or email to facilitate return of affected devices for replacement. Consignees with any questions can contact Fresenius Kabi by phone at 855-354-6387x1 or by email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology