Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1888-2025

Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter. Recall

Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1888-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 23, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
95,938 units
Distribution
See key details
FDA cause
Packaging

What is being recalled?

Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.

Affected model & UDI checker

Search the exact identifiers published for Z-1888-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8888123405PPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source noticeNot specified
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157788
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157825
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157856
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157894
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157719
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice10884521157750
Not specifiedPalindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.Check source notice20884521157924
FDA code information

Model No.8888123405P, 8888123411P, 8888128455P, 8888128461P, 8888119365P, 8888119371P, 8888133505P; UDI: 20884521157788, 20884521157825, 20884521157856, 20884521157894, 20884521157719, 10884521157750, 20884521157924; Lot/Serial No. 2204000102, 2228000068, 2228000070, 2234800148, 2234800159, 2234800160, 2302500128, 2305300136, 2315600104, 2315600113, 2315600119, 2315600122, 2318500135, 2318500136, 2318500140, 2320000255, 2332700275, 2335200414, 2412000238, 2412000239, 2414300228, 2414300233, 2418000150, 2426300169.

View the original FDA recall notice

Why was this product recalled?

Potential breach of sterile barrier packaging.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Mozarc Medical Us, LLC take?

An URGENT: MEDICAL DEVICE RECALL notification dated April 2025 was mailed to consignees on 4/23/25. This notification recommends that clinicians follow facility-specific policies and procedures for routine assessment of the hemodialysis access device for patency, function, and efficacy, as well as for monitoring patient status. Consignees are instructed to immediately quarantine and discontinue use of affected devices in inventory. Consignees are to complete the electronic Customer Confirmation form to facilitate obtaining a Return Goods Authorization number for recalled devices to be returned for credit or replaced. Product purchased through a distributor can be arranged through the distributor. Consignees with any questions about the Customer Confirmation form can contact IQVIA by phone at 1-256-680-9942 or by email at [email protected]. Consignees with questions about the recall can contact Medtronic Customer Care at 877-211-1850.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology