FoundationOne Companion Diagnostic (F1CDx) Recall
FoundationOne Companion Diagnostic (F1CDx)
FDA recall Class II
Ongoing
FDA lists this Class II recall for FoundationOne Companion Diagnostic (F1CDx). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1890-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 18, 2024
- FDA report date
- Jun 11, 2025
- Quantity in commerce
- 2 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
FoundationOne Companion Diagnostic (F1CDx)
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Report Numbers: ORD-XXXXX77-01, ORD-XXXXX91-01. Note: The above characters have been redacted to minimize patient information disclosure.
Why was this product recalled?
Two (2) reports were sent to customers with an incorrect negative claim on the claims page; the device variant information was displayed correctly in the tumor profiling section of both FDA-approved test reports.
What actions did Foundation Medicine, Inc. take?
On June 18, 2024, amended reports were sent to customers. It was to alert customers to an amended report that was recently released on 6/18/24. The new report replaces the report of 2/11/24. Additional contact was made on June 20, 2024 to communicate the error and provide the updated report to the customer.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1890-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology