Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033. Recall
Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1900-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 9, 2024
- FDA report date
- Jun 11, 2025
- Quantity in commerce
- 2 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Model/REF Number: 5881033. UDI-DI: B33158810330. Lot Number: M28188. Serial Numbers: 0107, 0108.
Why was this product recalled?
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
What actions did Silex Medical, LLC take?
Silex formally notified customers on 01/09/2024. Instructions to consignees included to inspect all instruments that they received from the affected lots for the defect. If any instruments had the issue, the customers were instructed to send it back to their facility under an RGA.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1900-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology