FDA lists this Class II recall for VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1903-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Apr 16, 2025
- FDA report date
- Jun 11, 2025
- Quantity in commerce
- 89 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test
Affected model & UDI checker
Search the exact identifiers published for Z-1903-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 20-006-021 | VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test | Check source notice | Not specified |
| Not specified | VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test | Check source notice | 00840487101599 |
FDA code information
UDI/DI 00840487101599, Lot Number 022525021A, exp. 08/26/2025
Why was this product recalled?
It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Luminex Corporation take?
Diasorin issued an Urgent Field Safety Notice to its consignees on 04/16/2025 via email. The notice explained the issue, potential risk to health, and requested the following: "Actions to be taken by the Customer/User 1. Please examine your inventory for the affected products, cease all use, and isolate or quarantine them pending disposal. 2. If you currently have any of the identified affected barcode serial numbers from VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A, please contact Technical Support at [email protected] to arrange for replacement and confirm the disposal of any remaining tests in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results for VERIGENE Gram-Negative Blood Culture Nucleic Acid Test (BC-GN). If you have encountered a false negative result with a patient sample tested with the potentially impacted portion of Lot 022525021A, please contact Technical Support at [email protected]. 5. If you are a distributor, provide a copy of this letter to any customers or organizations who may have received VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A. 6. Please complete the Acknowledgement & Receipt Form and return it to Technical Support. 7. For those with complaints, questions, or concerns please contact Technical Support at 1-877-785-2323 (U.S) or +1-512-381-4397 (Outside U.S.)."
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1903-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology