VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test Recall

VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test

FDA recall Class II

Completed

FDA lists this Class II recall for VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1903-2025
Recall classification
Class II
Enforcement status
Completed
Initiation date
Apr 16, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
89 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test

Affected model & UDI checker

Search the exact identifiers published for Z-1903-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
20-006-021VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic testCheck source noticeNot specified
Not specifiedVERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic testCheck source notice00840487101599
FDA code information

UDI/DI 00840487101599, Lot Number 022525021A, exp. 08/26/2025

View the original FDA recall notice

Why was this product recalled?

It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Luminex Corporation take?

Diasorin issued an Urgent Field Safety Notice to its consignees on 04/16/2025 via email. The notice explained the issue, potential risk to health, and requested the following: "Actions to be taken by the Customer/User 1. Please examine your inventory for the affected products, cease all use, and isolate or quarantine them pending disposal. 2. If you currently have any of the identified affected barcode serial numbers from VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A, please contact Technical Support at [email protected] to arrange for replacement and confirm the disposal of any remaining tests in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results for VERIGENE Gram-Negative Blood Culture Nucleic Acid Test (BC-GN). If you have encountered a false negative result with a patient sample tested with the potentially impacted portion of Lot 022525021A, please contact Technical Support at [email protected]. 5. If you are a distributor, provide a copy of this letter to any customers or organizations who may have received VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A. 6. Please complete the Acknowledgement & Receipt Form and return it to Technical Support. 7. For those with complaints, questions, or concerns please contact Technical Support at 1-877-785-2323 (U.S) or +1-512-381-4397 (Outside U.S.)."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology