Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1903-2026

OER-ELITE Endoscope Reprocessor Connecting Tubes. Recall

OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.

FDA recall Class II

Ongoing

FDA lists this Class II recall for OER-ELITE Endoscope Reprocessor Connecting Tubes.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1903-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 25, 2026
FDA report date
Apr 29, 2026
Quantity in commerce
2,089 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.

Affected model & UDI checker

Search the exact identifiers published for Z-1903-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MAJ-2115OER-ELITE Endoscope Reprocessor Connecting Tubes.Check source noticeNot specified
Not specifiedOER-ELITE Endoscope Reprocessor Connecting Tubes.Check source notice04953170404108
FDA code information

Model Number: MAJ-2115. UDI Number: 04953170404108. Lot Numbers: 09A, 0XA, 0YA, 0ZA, 11A, 12A, 25A, 26A, 27A, 28A, 29A, 2XA, 2YA, 2ZA, 31A, 32A, 33A, 34A, 35A, 36A, 37A, 38A, 39A, 3XA, 3YA, 3ZA, 41A, 42A, 43A, 44A, 45A, 46A, 47A, 48A, 49A, 4XA, 4YA, 51A.

View the original FDA recall notice

Why was this product recalled?

XXX

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 03/25/2026 via letter titled "URGENT: MEDICAL DEVICE REMOVAL." Consignees were notified of the issue and hazard involved and were instructed to examine inventory and OER-ELITE devices for the affected product. The current version has black lock levers. The previous versions have gray lock levers or integrated lock levers, which are the same color as the rest of the connector. Consignees were instructed that if they have sufficient quantities of the current version (black) to support reprocessing needs, stop using the previous version and set aside for return. They may proceed with using the current version, only. If they have insufficient quantities of the current version (black) available, and do not have another method of endoscope reprocessing readily available, they may continue using the previous version(s) consistent with the Instructions for Use, which require users to inspect the connecting tube prior to use. Transition to the current version as soon as possible by placing an order through Customer Service at 1-800-848-9024, option 2. When sufficient quantities of the current version (black) are obtained to support reprocessing needs, stop using the previous version and only use the current version going forward. Then arrange for return the previous version of the connector. Consignees were also requested to confirm receipt of the letter through the Olympus web portal and forward the notification to other users who may have the affected products if product were further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.