FDA lists this Class II recall for Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1907-2025
Recall classification
Class II
Enforcement status
Completed
Initiation date
Apr 30, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
10,176
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Device: REF/UDI-DI/Serial-Lot: AU480: B11810/14987666540664, B12183/14987666540671, B80091/16954701010015, B96692/14987666542965, B96693/14987666542972, C02654/14987666543283, C02655/14987666543290, C02845/16954701010022, C41919/14987666544174, C41920/14987666544181, N3659700/14987666535257, N3659800/14987666535264, N3659900/14987666535271, N3660000/14987666535288, N3660100/14987666535547, N3660200/14987666535554, N3660300/14987666535554, N3660400/14987666535769, N3660500/14987666536032, N3660600/14987666536049, N3660800/14987666536063, N3912400/14987666537084, N3912500/14987666537091, N3150700, N3151000, N3151300, N3151400, N3151500, N3151600, N3151900/ Serial: From 10155 to 2025021482. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66751 Kits containing MU993400, Lots: From 178713114 to 179433670

View the original FDA recall notice

Why was this product recalled?

A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.

What actions did Beckman Coulter Mishima K.K. take?

An "URGENT MEDICAL DEVICE RECALL" dated 4/24/2025 were mailed and emailed to customers on 4/30/2025 who were asked to do the following: - Check the analyzers, which meet the conditions below: a) Analyzers whose Serial No. is from 2023070718 to 2025021482. b) Analyzers whose sample probe has been replaced since July 1, 2023. - For analyzers that meet either of the above criteria, remove the sample probe from the instrument in accordance with IFU Section 6, Maintenance / As needed Maintenance/Replace a Sample or Reagent Probe. And then, check the groove on the head of the removed sample probe. a) If it has a marking line, continue to use. b) If it does not have a marking line, replace the sample probe with a probe not included in the impacted sample probe lot numbers. - Dispose impacted probes to ensure it is not used. Also check the sample probes included in the maintenance kits. Contact your firm representative for replacements. - If all the results of the actions above show only defective sample probes on your analyzers and in stock, stop measuring the assay items which sample volume is 2.0 microliters or less and contact your BEC representative for a replacement. - Firm recommends that customers should share the content of this letter with their laboratory and/or medical director to assess if a retrospective review of the patient results is needed. - Complete and return the response form via email to [email protected] -Complete and return the Replacement Order Form to [email protected] or fax to (866) 294-7850 OR Call Client Services at (800) 526-3821 If you have any questions regarding this notice, please contact the firm's Customer Support Center: From firm website: http://www.beckmancoulter.com For customers in the United States, if you need replacement product: Complete Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR Call Client Services at (800) 526-3821.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology