FDA RECALL RECORD · Z-1911-2025

Olympus EVIS EXERA III Duodenovideoscope. Recall

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus EVIS EXERA III Duodenovideoscope.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1911-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 5, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
12,195 units
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Affected model & UDI checker

Search the exact identifiers published for Z-1911-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
TJF-Q190VOlympus EVIS EXERA III Duodenovideoscope.Check source noticeNot specified
Not specifiedOlympus EVIS EXERA III Duodenovideoscope.Check source notice04953170405563
FDA code information

UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.

View the original FDA recall notice

Why was this product recalled?

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on 05/05/2025 via "URGENT Medical Device Correction" letter. Consignees were instructed to carefully review the notification, ensure all reprocessing and endoscopy unit personnel are completely knowledgeable and thoroughly trained on the updated reprocessing instructions, replace any copies of the reprocessing manual with the updated version (RC8415_02), complete and return the response form to recall web portal: https://olympusamerica.com/recall, enter the recall number: "0471" and complete the form as instructed, and forward the notification to affected customers if the affected product has been further distributed. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.