Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1914-2025

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Recall

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1914-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 28, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Process design

What is being recalled?

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.

Affected model & UDI checker

Search the exact identifiers published for Z-1914-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
84250320S.Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System.Check source notice07613327479515
84250320SLong Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System.Check source noticeNot specified
FDA code information

Catalog Number: 84250320S. GTIN: 07613327479515. Lot Number: K1BC9B7.

View the original FDA recall notice

Why was this product recalled?

Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.

What actions did STRYKER GmbH take?

Stryker notified consignees on about 05/28/2025 via FedEx letter. Consignees were instructed to review inventory and immediately quarantine any affected units on hand, and complete and return the Business Reply Form documenting receipt and product segregation. Upon receipt of the response form, Stryker will contact consignees to arrange for return of the affected units. If affected products have been further distributed, consignees were instructed to notify those customers and let Stryker know.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.