SCD Comfort Knee Length Small, SKU 84021; Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller Recall
SCD Comfort Knee Length Small, SKU 84021; Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller
FDA recall Class II
Completed
FDA lists this Class II recall for SCD Comfort Knee Length Small, SKU 84021; Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1917-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- May 13, 2025
- FDA report date
- Jun 18, 2025
- Quantity in commerce
- 6517 cases total
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
SCD Comfort Knee Length Small, SKU 84021; Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI-DI (GTIN) 10885380186516 (EA) 50885380186514 (BX) ALL LOTS
Why was this product recalled?
Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
What actions did Cardinal Health 200, LLC take?
On May 13, 2025, the firm began to notify customers via Urgent Medical Device Product Correction letters. Customers were informed of the issue, which exclusively impacts use of the Comfort Small Knee Sleeve. Other sleeve sizes are unaffected and operate as intended with the SmartFlow Controller. The product remains operational if the Comfort Small Knee Sleeve and SmartFlow Controller configuration is not utilized during therapy. A software update on the SmartFlow Controller is necessary before using the SmartFlow controller with the Comfort Small Knee sleeve configuration. Customers should discontinue use of the Comfort Small Knee Sleeve with the SmartFlow Controller and utilize a suitable alternative such as SM thigh or foot cuff. Customer service representatives will coordinate with customers to facilitate a software update which will allow the controllers to be utilized with the small knee sleeves without causing the error.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1917-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026