FDA RECALL RECORD · Z-1918-2025

Kendall SCD Comfort Sleeves Knee Length Small, Pediatric - For use with Kendall SCD SmartFlow, SKU 84021P Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller Recall

Kendall SCD Comfort Sleeves Knee Length Small, Pediatric - For use with Kendall SCD SmartFlow, SKU 84021P Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller

FDA recall Class II

Completed

FDA lists this Class II recall for Kendall SCD Comfort Sleeves Knee Length Small, Pediatric - For use with Kendall SCD SmartFlow, SKU 84021P Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1918-2025
Recall classification
Class II
Enforcement status
Completed
Initiation date
May 13, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
6517 cases total
Distribution
See key details
FDA cause
Software change control

What is being recalled?

Kendall SCD Comfort Sleeves Knee Length Small, Pediatric - For use with Kendall SCD SmartFlow, SKU 84021P Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI (GTIN) 10885380197925 (EA) 50885380197923 (BX) ALL LOTS

View the original FDA recall notice

Why was this product recalled?

Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.

What actions did Cardinal Health 200, LLC take?

On May 13, 2025, the firm began to notify customers via Urgent Medical Device Product Correction letters. Customers were informed of the issue, which exclusively impacts use of the Comfort Small Knee Sleeve. Other sleeve sizes are unaffected and operate as intended with the SmartFlow Controller. The product remains operational if the Comfort Small Knee Sleeve and SmartFlow Controller configuration is not utilized during therapy. A software update on the SmartFlow Controller is necessary before using the SmartFlow controller with the Comfort Small Knee sleeve configuration. Customers should discontinue use of the Comfort Small Knee Sleeve with the SmartFlow Controller and utilize a suitable alternative such as SM thigh or foot cuff. Customer service representatives will coordinate with customers to facilitate a software update which will allow the controllers to be utilized with the small knee sleeves without causing the error.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology