Access Total T4 Calibrator, Catalog No. 33805 Recall

Access Total T4 Calibrator, Catalog No. 33805

FDA recall Class II

Ongoing

FDA lists this Class II recall for Access Total T4 Calibrator, Catalog No. 33805. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1922-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 16, 2026
FDA report date
Apr 29, 2026
Quantity in commerce
2068 US, 2830 OUS
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Access Total T4 Calibrator, Catalog No. 33805

Affected model & UDI checker

Search the exact identifiers published for Z-1922-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
33805Access Total T4 Calibrator, Catalog No. 33805Check source noticeNot specified
Not specifiedAccess Total T4 Calibrator, Catalog No. 33805Check source notice15099590225872
FDA code information

UDI-DI 15099590225872 Lots 538729, 538815, 538898, 539107, 539243

View the original FDA recall notice

Why was this product recalled?

Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.

What actions did BECKMAN COULTER, INC. take?

Pending

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.