Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1926-2026

Flamingo Funnel Large, Model Number SQ20012-03 Recall

Flamingo Funnel Large, Model Number SQ20012-03

FDA recall Class II

Ongoing

FDA lists this Class II recall for Flamingo Funnel Large, Model Number SQ20012-03. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1926-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 16, 2026
FDA report date
Apr 29, 2026
Quantity in commerce
2505 units total
Distribution
See key details
FDA cause
Packaging change control

What is being recalled?

Flamingo Funnel Large, Model Number SQ20012-03

Affected model & UDI checker

Search the exact identifiers published for Z-1926-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SQ20012-03Flamingo Funnel Large, Model Number SQ20012-03Check source noticeNot specified
FDA code information

UDI-DI: 408882777851966; Lot 8264501

View the original FDA recall notice

Why was this product recalled?

Certain lots of product were associated with a manufacturing nonconformance involving use of a non-gamma sterilization compatible drape.

What actions did SurgiSmoke Solutions take?

An URGENT: MEDICAL DEVICE RECALL notification letter was sent to customers beginning 3/16/26. Actions to be taken by the Customer/End User: 1. Immediately discontinue use of this product. Note: The lots listed in this notice are the only lots impacted. All subsequently built product contains a correct, gamma compatible drape. 2. Dispose of impacted product according to local and national waste disposal procedures. Note: ALL PRODUCT MUST BE ACCOUNTED FOR ON THE MEDICAL DEVICE RECALL RETURN RESPONSE ACKNOWLEDGEMENT AND RECEIPT FORM. 3. Acknowledge receipt of this notice. Complete and return the attached Medical Device Recall Return Response Acknowledgement and Receipt Form via phone, US Mail, or email even if you no longer have any affected product in inventory: Phone: Communicate the information required in the attached form by calling the toll free number: 1-800-557-3698 - available M-F between 8:00AM and 6:00PM CST US Mail: Complete the attached form, and mail it back using the enclosed self-addressed, stamped envelope (if notice received by mail) Email: Complete the attached form and return it to [email protected] Please forward this notification to any additional customers to whom the affected product may have been distributed. Type of Action by the Company: In addition to this Recall activity, Nascent Surgical has initiated Corrective Action and Preventive Action measures that include but are not limited to: Correcting the Bill of Materials to include a gamma compatible drape (completed Jan 2025) Changing the packaging configuration to remove the fold (completed Jan 2025) Improving Quality System Planning and other procedures to prevent recurrence Other Information: If you have additional questions about this notice, please contact us: By phone 1-800-557-3698 (toll free) - available M-F between 8:00AM and 6:00PM CST By emailing [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology