EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G. Recall
EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE¿ device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G.
FDA recall Class II
Ongoing
FDA lists this Class II recall for EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1927-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 23, 2025
- FDA report date
- Jun 18, 2025
- Quantity in commerce
- 4 units in U.S.
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G.
Affected model & UDI checker
Search the exact identifiers published for Z-1927-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 30-0001 | EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G. | Check source notice | Not specified |
| Not specified | EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G. | Check source notice | 03760305400031 |
FDA code information
Model Number: 30-0001 UDI-DI code: 03760305400031 Software Version: 1.0.6.1 Serial Numbers (U.S.): 0322002, 0323006, 0323003 and 0324001
Why was this product recalled?
Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user.
What actions did QUANTUM SURGICAL SAS take?
On or about 05/23/2025, the firm sent via emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due to a software issue the EPIONE device can place the instrument guide at an incorrect position, compared to the planning defined by the user. Customers are instructed to: 1. Pending the implementation of a corrective action, Quantum Surgical asks all EPIONE device users to NOT plan trajectories along a strictly horizontal direction. To determine if a trajectory is strictly horizontal the following methods can be used: . Patient images acquired in prone or supine orientation: o If the entry and target point are on the same native (non-oblique) axial plane and are on the same native (non-oblique) coronal plane -> the trajectory is strictly horizontal. o If the trajectory is perpendicular to the sagittal plane -> the trajectory is strictly horizontal. . Patient images acquired in lateral decubitus orientation: o If the entry and target point are on the same native (non-oblique) axial plane and are on the same native (non-oblique) sagittal plane -> the trajectory is strictly horizontal. o If the trajectory is perpendicular to the coronal plane -> the trajectory is strictly horizontal. Note - Quantum Surgical reminds that the instrument insertion shall be initiated only after the "Robot on Target" message is displayed, as mentioned in Epione User Manual on page 100. For question or assistance, contact Quantum Surgical on +33 (0) 4 48 19 40 50 between 9am and 6pm, Monday to Friday.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1927-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology