FDA RECALL RECORD · Z-1942-2026

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000 Recall

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

FDA recall Class I

Ongoing

FDA lists this Class I recall for VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1942-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 25, 2026
FDA report date
May 13, 2026
Quantity in commerce
42
Distribution
See key details
FDA cause
Process change control

What is being recalled?

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Affected model & UDI checker

Search the exact identifiers published for Z-1942-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PRT-01198-000VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000Check source noticeNot specified
PRT-01185-000VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000Check source noticeNot specified
FDA code information

UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532

View the original FDA recall notice

Why was this product recalled?

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

What actions did Ventec Life Systems, Inc. take?

On 3/25/2026 recall notices were mailed to customers who were asked to do the following: 1. Immediately discontinue use of affected devices. 2. Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3. Use a non-affected ventilator for patient support. 4. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5. Complete and return the response form via email to [email protected] 6. If you are no longer the owner of any of these devices, you need to complete an ownership transfer form at: VentecLife.com/tracking so firm can follow up with the current owner of the impacted device(s). If you have questions, you can contact the firm at 844-698-6276

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history