Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2) Recall
Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)
FDA recall Class I
Ongoing
FDA lists this Class I recall for Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1946-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2025
- FDA report date
- Jun 25, 2025
- Quantity in commerce
- 38,127 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
All Lots/ UDI: STK-DO-006 -00386270003188 STK-DO-013 -00386270002327 STK-DO-103-00386270002969 STK-DO-109 -00386270002792
Why was this product recalled?
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
What actions did Dexcom, Inc. take?
On May 12, 2025, Dexcom issued a "Urgent: Medical Device Correction" Notification to affected consignees via E-Mail. On June 9, 2025, Dexcom issued an Urgent: Medical Device Correction update Dexcom asked consignees to take the following actions: 1. Verify if your receiver is affected by this issue: Check the serial number for any receiver(s) you use: https://dexcom.com/en-us/checkreceiver 2. If your receiver is affected, contact Dexcom Technical Support at 1-844-478-1600 to coordinate return of the affected receiver and a no charge replacement. 3. Whether or not your receiver is affected by this notification, Dexcom continues to recommend regular testing of the speaker. Test speaker function every time you charge your receiver. When you charge, a reminder prompt will invite you to test speaker function. 4.If you need assistance related to this issue or notification, please call Dexcom Technical Support at 1- 844-478-1600. Technical Support is available 24 hours a day/7 days a week to assist you.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1946-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology