Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kit Recall

Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kit. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1948-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 24, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
952 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component

Affected model & UDI checker

Search the exact identifiers published for Z-1948-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
M308Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kitCheck source noticeNot specified
Not specifiedBrand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kitCheck source notice30014613336167
FDA code information

Lot Code: Catalog number: M308 UDI-DI: 30014613336167; Lots: 245251 - Exp. 2025-Oct-25 246941 - Exp. 2025-Dec-03 250163 - Exp. 2026-Feb-06 259686 - Exp. 2026-Sep-03 261072 - Exp. 2026-Oct-02 262113 - Exp. 2026-Oct-23 263331 - Exp. 2026-Nov-22 264065 - Exp. 2026-Dec-11 265007 - Exp. 2027-Jan-03 (corrected exp date from 2026-Dec-31) 266928 - Exp. 2026-Dec-31 Additional Lots: 265007-1 - Exp. 2026-Dec-31 268479 and above - Various Expiration dates *Until further notice, all future lots, both supplied by QuidelOrtho or supplied by its distributors, will be affected.

View the original FDA recall notice

Why was this product recalled?

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?

On May 6, 2025, URGENT PRODUCT CORRECTION NOTIFICATION letters to customers. REQUIRED ACTION " Do not report Bordetella parapertussis (BPP) positive results when testing with the affected lot numbers. " Use the Solana Bordetella Complete Assay only for reporting Bordetella pertussis (BP) results and negative Bordetella parapertussis (BPP) results only when using the affected lots, as they are not affected. " Review the content of this communication with your Medical Director and retain this letter for your laboratory. " Complete the enclosed Confirmation of Receipt form no later than May 14, 2025. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-874-1517 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology