Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1950-2026

Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Recall

Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery. The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges. Component: Not Applicable

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1950-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 23, 2026
FDA report date
May 6, 2026
Quantity in commerce
45173
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery. The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges. Component: Not Applicable

Affected model & UDI checker

Search the exact identifiers published for Z-1950-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CI-5501-110Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery PakCheck source noticeNot specified
FDA code information

Lot Code: All lots manufactured up to February 3, 2026. Refer to Exhibit_0006

View the original FDA recall notice

Why was this product recalled?

The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.

What actions did Advanced Bionics, LLC take?

On 03/24/2026, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL M ZN-AIR BATTERY PAK (CI-5501)" Letter to inform them that, Advanced Bionics is voluntarily recalling the non-tamperproof battery door from the packaged configuration for all M Zn-Air Battery Pak (CI-5501), due to the non-tamperproof battery door does not require the use of a tool to replace the button batteries and therefore could allow unintended access to the button cell batteries within the battery compartment. Customer are instructed to: 1. Identify the Non-Tamperproof Battery Door - confirm which door is installed by performing the following check: -When inserted into the battery cartridge, the non-tamperproof door can be removed without using a tool. -The tamperproof battery door requires a tool for removal once installed. If you must use a tool to remove the door, you have the tamperproof door installed. 2. Remove and Dispose of the Non-Tamperproof Battery Door - Once identified, ensure the non-tamperproof battery door is removed from the battery cartridge. Do not reinstall or use this component. 3. Complete the Acknowledgement & Disposal Form (required) - After disposing of the non-tamperproof battery door, confirm that the component has been removed from service and disposed of appropriately by completing the Acknowledgement & Disposal Form linked in your email. For questions - contact customer service at 1-877-829-0026 or email [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology