FDA RECALL RECORD · Z-1951-2026

Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296. Recall

Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1951-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 14, 2026
FDA report date
May 6, 2026
Quantity in commerce
3 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.

Affected model & UDI checker

Search the exact identifiers published for Z-1951-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPhilips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.Check source notice00884838004108
FDA code information

1. Model Number (REF): 781196. UDI: None. Serial Numbers: 21957. 2. Model Number (REF): 781296. UDI: (01)00884838004108(21)32039, (01)00884838004108(21)32095. Serial Numbers: 32039, 32095.

View the original FDA recall notice

Why was this product recalled?

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America take?

Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.