Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1977-2025

REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile Recall

REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile

FDA recall Class II

Ongoing

FDA lists this Class II recall for REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1977-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 24, 2024
FDA report date
Jun 25, 2025
Quantity in commerce
2 units
Distribution
See key details
FDA cause
Use error

What is being recalled?

REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile

Affected model & UDI checker

Search the exact identifiers published for Z-1977-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
99-709245REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, SterileCheck source noticeNot specified
D9mmREF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, SterileCheck source noticeNot specified
Not specifiedREF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, SterileCheck source notice18059015373653
FDA code information

Lot # B4771336 and B4771337, UDI: 18059015373653

View the original FDA recall notice

Why was this product recalled?

Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ORTHOFIX SRL take?

No official notification was sent to customers. Orthofix representative was present during the first clinical case and both devices were returned as part of the complaint investigation process.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology