Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1981-2026

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ Recall

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

FDA recall Class I

Ongoing

FDA lists this Class I recall for Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1981-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 19, 2026
FDA report date
May 13, 2026
Quantity in commerce
7,311 kits
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Affected model & UDI checker

Search the exact identifiers published for Z-1981-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ANCA80APConvenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQCheck source noticeNot specified
Not specifiedConvenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQCheck source notice191072232168
Not specifiedConvenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQCheck source notice191072236678
FDA code information

ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911

View the original FDA recall notice

Why was this product recalled?

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AMERICAN CONTRACT SYSTEMS, INC. take?

Firm notified distributors with an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" dated March 19, 2026, via email. Distributors were instructed to notify end-users. An updated letter with additional kits was sent to the customer on 3/20/26. Customers should quarantine all affected product. Customers may add warning labels to affected ACS kits to ensure all users are aware of prescribed actions that were identified in the Medline recall notice. The warning labels state, "Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure." Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, contact the Regulatory Affairs Manager, by email: [email protected] or at 419-455-2090.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology