STA - Owren Koller. Recall
STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.
FDA recall Class II
Ongoing
FDA lists this Class II recall for STA - Owren Koller.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1993-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2025
- FDA report date
- Jun 25, 2025
- Quantity in commerce
- 939 US ( 2,810 total worldwide)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.
Affected model & UDI checker
Search the exact identifiers published for Z-1993-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 00360 | STA - Owren Koller. | Check source notice | Not specified |
| Not specified | STA - Owren Koller. | Check source notice | 13607450003601 |
FDA code information
UDI: (01)13607450003601(11)240331(17)260331(10)270902(241)00360. Lot Number: 270902.
Why was this product recalled?
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Diagnostica Stago, Inc. take?
Diagnostica Stago, Inc. notified consignees on 05/21/2025 via email. Consignees were instructed to review inventory for any affected units, stop using the affected lot and destroy it, and complete and return the acknowledgement form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1993-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology