Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system Recall
Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
FDA recall Class II
Ongoing
FDA lists this Class II recall for Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1994-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 9, 2025
- FDA report date
- Jul 2, 2025
- Quantity in commerce
- 223
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726; K9303300/05415334002180/2531537815, 2531557845, 2531557866, 2531558588, 2531558727, 2531558738, 2531558753, 2531558754, 2531558759, 2531561761, 2531561785, 2531572541, 2531572557, 2531572568, 2531572576, 2531572847, 2531572862, 2531572872, 2531633606, 2531633662; K9303311/05415334026810/2531537674, 2531572141, 2531580810, 2531580935, 2531582319, 2531582331, 2531582341, 2531588386, 2531588401, 2531588404, 2531588456, 2531591560, 2531592036, 2531592037, 2531592046, 2531593872, 2531594457, 2531594475, 2531594481, 2531594503, 2531594515, 2531597664, 2531597713, 2531597717, 2531597726, 2531597795, 2531597806, 2531597809, 2531597810, 2531597820, 2531599836, 2531605356, 2531605390; K9306210/2531542260, 2531542263, 2531542270, 2531542292, 2531543553, 2531543568, 2531543588, 2531582676, 2531582691, 2531588558, 2531588562, 2531593741, 2531593749, 2531593752, 2531593763, 2531593764, 2531593768, 2531593777, 2531593784, 2531595829, 2531595831, 2531601786; K9306220/ 2531540107, 2531542441, 2531542470, 2531543538, 2531543549, 2531543611, 2531543613, 2531543640, 2531563132, 2531572740, 2531572759, 2531572770, 2531572814, 2531572827, 2531575652, 2531594331, 2531594332, 2531594336, 2531594348, 2531594350; K9306410/2531572228, 2531582713, 2531582724, 2531582725, 2531584168, 2531584176, 2531584191, 2531588591, 2531588593, 2531588598, 2531588616, 2531588626, 2531604519, 2531604535, 2531604544, 2531604545, 2531604549, 2531604550, 2531604551, 2531604593, 2531604637, 2531604649, 2531604658; K9306420/ 2531537844, 2531537870, 2531537889, 2531537901, 2531537902, 2531556037, 2531556043, 2531572419, 2531572433, 2531572440, 2531595759
Why was this product recalled?
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
What actions did Barco N.V. take?
On 5/9/2025, recall notices started being distributed to customers informing them of the following: 1) Assess & locate the non-installed affected units and contain them from further distribution. B) Assess & locate affected units that are installed at your customer side. Please make your customers aware of this notification. C) The impacted units will need to be sent back. Further information on the swaps will be provided in the coming days. D) Fill in and return the Acknowledgement Letter. If you don t have the Key Account Manager s contact details, or need other support, you can contact the firm 24/7 via their Support web page: https://www.barco.com/en/support
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1994-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology