FDA lists this Class II recall for Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1995-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 27, 2026
FDA report date
May 6, 2026
Quantity in commerce
6 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments

Affected model & UDI checker

Search the exact identifiers published for Z-1995-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedStraumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant AbutmentsCheck source notice07630031759764
FDA code information

Article 010.0158. UDI: (01)07630031759764(11)YYMMDD(10)XXXXX. Project Numbers: 28000US_00622_00_b00, 28000US_00431_00_b00, 28000US_00489_00_b00, 28000US_00792_00_b00, 28000US_00908_00_b00, 28000US_00908_01_b00.

View the original FDA recall notice

Why was this product recalled?

Includes an incorrect screw seat interface.

What actions did Straumann USA LLC take?

Straumann USA notified consignees on about 03/27/2026 via telephone call, mailed letter, or email. Consignees were instructed to examine inventory for affected product, cease use and quarantine any affected units on hand, arrange for the return of all affected units and complete and return the provided Customer Confirmation Form. Consignees were also instructed to no contact Straumann contact Jennifer Jackson to organize a remake of the units, if any were used. Once the new restoration is available, schedule the patient to replace the restoration.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.