iLet Bionic Pancreas, REF: BB1001 Recall
iLet Bionic Pancreas, REF: BB1001
FDA recall Class II
Ongoing
FDA lists this Class II recall for iLet Bionic Pancreas, REF: BB1001. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1996-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 12, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 1,080
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
iLet Bionic Pancreas, REF: BB1001
Affected model & UDI checker
Search the exact identifiers published for Z-1996-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| BB1001 | iLet Bionic Pancreas, REF: BB1001 | Check source notice | Not specified |
FDA code information
Software versions 1.4.3, and 1.4.2
Why was this product recalled?
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Beta Bionics, Inc. take?
On 3/12/2026, correction notices were emailed to customers who were asked to do the following: Update device software to host software 1.4.4 and up. Update your software 1. Download or update your iLet app on your phone. Read the iLet App User Guide: https://www.betabionics.com/wp-content/uploads/LA000056_F-iLet-Mobile-App-User-Guide.pdf 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. -Distributors are asked to provide patient contact information so the firm can provide the correction notice to these patients. - Complete the acknowledgment survey https://sprw.io/stt-6HnQU If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1996-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.