Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1998-2026

Heraeus, PALAMIX duo. Material Number: 66057897. Recall

Heraeus, PALAMIX duo. Material Number: 66057897.

FDA recall Class III

Ongoing

FDA lists this Class III recall for Heraeus, PALAMIX duo. Material Number: 66057897.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1998-2026
Recall classification
Class III
Enforcement status
Ongoing
Initiation date
Mar 25, 2026
FDA report date
May 6, 2026
Quantity in commerce
1,125 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Heraeus, PALAMIX duo. Material Number: 66057897.

Affected model & UDI checker

Search the exact identifiers published for Z-1998-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedHeraeus, PALAMIX duo. Material Number: 66057897.Check source notice04260102133617
FDA code information

Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;

View the original FDA recall notice

Why was this product recalled?

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What actions did Heraeus Medical GmbH (Dental Division) take?

On 03/25/2026 Field Safety Notice letters were sent to customers. Actions to be taken: 1. Read this Field Safety Notice and ensure that all relevant hospital departments are informed about its content. 2. Please complete the attached customer reply form and send it to [email protected] within five (5) calendar days. This form must be returned even if you no longer use the device. 3. Keep a copy of the attached customer reply form with your records in the event of a compliance audit of your facility.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology