Brand Name: Leksell Vantage Arc System REF: 1053958 Recall
Brand Name: Leksell Vantage Arc System REF: 1053958
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Leksell Vantage Arc System REF: 1053958. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2006-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 19, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 768 (364Left/364Right)
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Leksell Vantage Arc System REF: 1053958
Affected model & UDI checker
Search the exact identifiers published for Z-2006-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1053958 | Brand Name: Leksell Vantage Arc System REF: 1053958 | Check source notice | Not specified |
| Not specified | Brand Name: Leksell Vantage Arc System REF: 1053958 | Check source notice | 07340048304856 |
| Not specified | Brand Name: Leksell Vantage Arc System REF: 1053958 | Check source notice | 07340048304818 |
| Not specified | Brand Name: Leksell Vantage Arc System REF: 1053958 | Check source notice | 07340048304801 |
FDA code information
Lot Code: Arc System: UDI:07340048304856 Parts of Arc System: Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625 Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675
Why was this product recalled?
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
What actions did Elekta, Inc. take?
On March 19, 2026 Elekta issued a "Urgent: Medical Device Recall" notification via e-mail. Elekta asked consignees 1. Before each operation, products should be cleaned, sterilized (unless specified otherwise), and checked for proper functionality. 2. Make sure to examine all scales and angles to make sure that they are correctly set and securely fixed before proceeding. 3. Make sure to obtain a second person in the OR room to read the coordinate settings so as to ensure they are correct and check they are securely locked. 4. Before you continue, check that all parts are assembled correctly, firmly attached, and locked in place. 5. Inspect the equipment for visible signs of damage before clinical use. Do not use damaged equipment for any purpose. If you suspect that any part is damaged, contact your Elekta representative immediately. 6. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 7. Advise the appropriate personnel working with this product, on the content of this letter. If the product has been transferred to another facility, please provide them with a copy of this communication.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2006-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology