FDA RECALL RECORD · Z-2007-2026

Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case Recall

Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case

FDA recall Class I

Ongoing

FDA lists this Class I recall for Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2007-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 17, 2026
FDA report date
May 13, 2026
Quantity in commerce
2030 kits
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case

Affected model & UDI checker

Search the exact identifiers published for Z-2007-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BHCA49KMedical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/CaseCheck source noticeNot specified
Not specifiedMedical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/CaseCheck source notice20809160490294
FDA code information

UDI:20809160490294/ Lot numbers: 342509, 346066, 338434, 339423, 337365, 335489, 336189, 336789

View the original FDA recall notice

Why was this product recalled?

Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"

What actions did Medical Action Industries, Inc. 306 take?

On April 22, 2026, Medial Action Industries, Inc. (subsidiary of Owens & Minor) issued a "Urgent: Medical Device Recall Notification Update" via E-Mail to affected consignees. The updated notification provided clarity and and additional information for the recall strategy. Initial notification was issued on March 17, 2026. Medical Action Industries asked consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product. Convenience kits can still be used once properly labeled. 2. Add warning labels to affected Medical Action kits to ensure all users are aware of the actions below and identified in the Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. 3. If your firm used the labeling template provided with the initial letter, please adhere the revised label over the existing template label. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by April 30, 2026. If you have any questions, contact Reg. Affairs Manager, Kitting, Global Products, by email: [email protected] or 470-280-4277.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology