Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2016-2025

Omnipod 5 Automated Insulin Delivery System, iOS Application. Recall

Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface that controls the Omnipod 5 Automated Insulin Delivery System. It is used to activate/deactivate Pods, display alerts/alarms, and send insulin delivery commands for execution to the Pod. Only available in the US.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Omnipod 5 Automated Insulin Delivery System, iOS Application.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2016-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 23, 2025
FDA report date
Jul 2, 2025
Quantity in commerce
122,823 downloads
Distribution
See key details
FDA cause
Software design

What is being recalled?

Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface that controls the Omnipod 5 Automated Insulin Delivery System. It is used to activate/deactivate Pods, display alerts/alarms, and send insulin delivery commands for execution to the Pod. Only available in the US.

Affected model & UDI checker

Search the exact identifiers published for Z-2016-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PT-000664/M009-S-APOmnipod 5 Automated Insulin Delivery System, iOS Application.Check source noticeNot specified
Not specifiedOmnipod 5 Automated Insulin Delivery System, iOS Application.Check source notice10385083000336
FDA code information

Model Number: PT-000664/M009-S-AP. UDI-DI: 10385083000336. All software versions 1.X.X

View the original FDA recall notice

Why was this product recalled?

Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Insulet Corporation take?

Insulet notified health care professionals and users on 05/23/2025 via emailed letter. Consignees were instructed to make sure your phone s keyboard and region settings match the approved country for use currently, the U.S. only and this issue does not affect the Insulet-provided Omnipod 5 Controllers or the Omnipod 5 App on compatible Android phones. A software update will be pushed to all devices in the coming weeks, and each consignee will be notified via email and in-app notification on the device when it is ready to install. The fix will allow commas to be recognized as decimal points when using the Omnipod 5 App for iPhone. The email notification to users included a section to formally acknowledge the notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology