PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile Recall
PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile
FDA recall Class II
Ongoing
FDA lists this Class II recall for PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2025-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 19, 2025
- FDA report date
- Jul 2, 2025
- Quantity in commerce
- 9 units
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile
Affected model & UDI checker
Search the exact identifiers published for Z-2025-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1367.15.704 | PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile | Check source notice | Not specified |
| Not specified | PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile | Check source notice | 08033390201834 |
FDA code information
Lot number 2421400/UDI: 08033390201834
Why was this product recalled?
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Limacorporate S.p.A. take?
On May 19, 2025, Enovis (Limacorporate S.p.A) issued a recall notification via e-mail. Enovis provided customers with an updated notification on May 28, 2025 to provide additional information. Enovis asked consignees to take the following steps: 1. Review your stock for the products and lot numbers for the items to be returned and quarantine immediately the affected devices. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. If you have further distributed the product, please provide this recall notification to those consignees; have them complete the acknowledgement form; and notify anyone that they have further distributed the affected product to. 3. Complete the attached acknowledgement form and contact Customer Service at 1-800-456-8696 or through your regional email address, to place a replacement request. You will be provided with an RMA number and shipping label to return the affected product. 4. Return all devices using the RMA number and shipping label provided by Customer Service. Devices must be sent back to: Lima U.S.A., Inc. | 2001 NE Green Oaks Blvd. Ste.100 | Arlington, Texas 76006 | USA within 15 days together with a signed hard copy of the following acknowledgement form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2025-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology