Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2027-2025

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300; Recall

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2027-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 22, 2025
FDA report date
Jul 2, 2025
Quantity in commerce
1,625 units (1450 US, 175 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

Affected model & UDI checker

Search the exact identifiers published for Z-2027-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedProduct: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;Check source notice00630414595771
FDA code information

Siemens Material Number (SMN): 11099300; UDI-DI: 00630414595771; Lot Number: 0000195412;

View the original FDA recall notice

Why was this product recalled?

Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

What actions did Siemens Healthcare Diagnostics, Inc. take?

On May 22, 2025 URGENT FIELD SAFETY NOTICE letters were sent to OUS customers and URGENT MEDICAL DEVICE CORRECTION letters were sent to US customers. Customer Actions Please discontinue use of and discard all inventory of the diluent lot number listed in the table above (Products Section). If lot number 0000195412 is currently on your analyzer, remove the bottle and replace with an alternate lot number. See Appendix for detailed instructions. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.