Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226. Recall

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2032-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 30, 2025
FDA report date
Jul 16, 2025
Quantity in commerce
9839 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025

View the original FDA recall notice

Why was this product recalled?

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Baxter Healthcare Corporation take?

Baxter issued an URGENT PRODUCT RECALL notice to is consignees on 05/30/2025 via USPS first class mail. The notice described the problem with the device, hazard, and requested the following: "Actions to be taken by customers 1. Please locate and stop using all Q-link 13 components on any of the products listed in the affected product table. 2. Contact Baxter Technical Support to discuss alternative options, if necessary, at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at [email protected]. 3. Please post this letter in areas where affected mobile lifts are stored and used. 4. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 6. Please forward a copy of this communication to all departments within your institution that use the affected mobile lifts. 7. If you are a dealer, wholesaler or distributor/reseller that distributed any affected product to other facilities, please forward this communication to your customers and check the associated box on the customer portal. Further Information and Support For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 Option 2 between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Sup

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology