Brand Name: VITROS Product Name: VITROS 5600 Integrated System Recall
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: VITROS Product Name: VITROS 5600 Integrated System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2038-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 9, 2025
- FDA report date
- Jul 9, 2025
- Quantity in commerce
- 3117 units (813 US, 2304 OUS)
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
Affected model & UDI checker
Search the exact identifiers published for Z-2038-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6802413 | Brand Name: VITROS Product Name: VITROS 5600 Integrated System | Check source notice | 10758750002740 |
| 6802915 | Brand Name: VITROS Product Name: VITROS 5600 Integrated System | Check source notice | 10758750002740 |
FDA code information
VITROS 5600 Integrated System; Product Code: 6802413; UDI: 10758750002740; VITROS 5600 Integrated System - Certified/Refurbished; Product Code: 6802915; UDI: 10758750002740; Range of installed serial numbers: 56000143-56005096 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802915 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process.
Why was this product recalled?
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?
On May 22, 2025, "URGENT PRODUCT CORRECTION NOTIFICATION" letters were sent to customers. REQUIRED ACTIONS - To prevent this issue from occurring, the shelf expiration date must be entered when manually loading VITROS Diluent Packs on VITROS 4600/5600/XT 7600 Systems. - View the Reagent Management Supply 3 screen to check for any onboard VITROS Diluent Packs marked as expired. o For all VITROS Diluent Packs marked as expired, navigate to View By Reagent View Dil/ANC and confirm the shelf expiration date matches the shelf expiration date on the product labeling. o If the shelf expiration dates do not match, discard the VITROS Diluent Pack and load a new VITROS Diluent Pack. QuidelOrtho will credit your discarded inventory. (Credit will only be issued for within-expiry products.) - Complete and return the enclosed Confirmation of Receipt form no later than May 30, 2025. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. - Save this notification with your User Documentation or post this notification by each VITROS 4600/5600/XT 7600 System in your laboratory until the issue has been resolved. - Please forward this notification if the affected product was distributed outside of your facility. - If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Resolution QuidelOrtho has determined root cause, and this issue will be resolved in an upcoming software update. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2038-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology