Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2040-2026

Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317. Recall

Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2040-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 16, 2026
FDA report date
May 6, 2026
Quantity in commerce
4050 kits
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Medline Kit SKU DYNDH1520: UDI/DI each 10193489197679, UDI/DI case 20193489197676, Lot Number : 2025101301; Medline Kit SKU DYNDH1520: UDI/DI each 10193489197679, UDI/DI case 20193489197676, Lot Number : 2025113001; Medline Kit SKU EBSI1317: UDI/DI each 10193489195804, UDI/DI case 20193489195801, Lot Number : 2025120390; Medline Kit SKU EBSI1317: UDI/DI each 10193489195804, UDI/DI case 20193489195801, Lot Number : 2026022390.

View the original FDA recall notice

Why was this product recalled?

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

What actions did Medline Industries, LP take?

Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/16/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the affected product be quarantined pending additional labeling. Consignees were directed to respond to the notice via Website link: https://recalls.medline.com, and upon receipt of the submitted response form, the account will receive over-labels to place on affected inventory, with instructions for the user to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight. Distributors or those who have further distributed the product are directed to notify those receiving the affected kits, and to include the customers quantities on the response form. Upon receipt of your submitted response form, you will receive over-labels to provide to the customers to affix to their affected product. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.