Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2041-2025

HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323 Recall

HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323

FDA recall Class II

Ongoing

FDA lists this Class II recall for HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2041-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 6, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
4300 ea
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323

Affected model & UDI checker

Search the exact identifiers published for Z-2041-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
G323HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323Check source noticeNot specified
FDA code information

UDI-DI 008165760257704 G323 lot 653594P: Manufactured on March 6, 2025; Expires on May 15, 2025 G323 lot 655101P: Manufactured on April 4, 2025; Expires on June 13, 2025 G323 lot 656816P: Manufactured on April 16, 2025; Expires on June 25, 2025

View the original FDA recall notice

Why was this product recalled?

Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae.

What actions did Hardy Diagnostics take?

On May 6, 2025, the firm began notifying customers via "URGENT Medical Device Recall Notification" letters. Customers were instructed to discard any remaining stock and to destroy it according to local regulations. If product was further distributed, notify the consignees who received the product. Please retain this letter or post the letter where the product is stored to ensure the users are aware of this defect. For replacements or credit, please contact the firm's Technical Services Department at (800) 266-2222, option 2 or via email at [email protected]. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to your invoice upon request. If you have any questions, call 805-346-2766 or email: [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.