Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Recall
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2046-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 2, 2026
- FDA report date
- May 13, 2026
- Quantity in commerce
- 2781
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Affected model & UDI checker
Search the exact identifiers published for Z-2046-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| IS4000 | Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System | Check source notice | Not specified |
| Not specified | Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System | Check source notice | 00886874114216 |
FDA code information
Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI: 00886874114216 FDA Medical Device Listing Number: D295931 See attachment Affected Product List
Why was this product recalled?
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
What actions did Intuitive Surgical, Inc. take?
On 04/02/2026, the firm sent via 2-day mailing service an "Urgent Medical Device Correction" letter to customers informing them that Intuitive identified that the screws on the cannula mount of the USM from a specific population are susceptible to breaking over time. When the cannula mount experiences a breakage of all six screws, the mount can separate from the rest of the USM. Breakage of any fewer than six screws will not result in cannula mount separation, or impact to clinical performance. How to recognize issue/failure: Screw breakage or cannula mount loosening may be observable as visible or tactile separation of the cannula mount from the body of the USM. Prior to a procedure, closely inspect the joint between the cannula mount and insertion axis that is joined by the screws in Figure 1 in the Urgent Medical Device Correction Letter. Manipulate the cannula mount joint by hand to ensure there is no movement or looseness. If looseness or separation is observed, discontinue use of the USM and contact representative to schedule a site visit. If all six screws were to break intra-operatively, the user may also observe imprecise or unintended instrument tip motion. If observed imprecise or unintended instrument tip motion, inspect the cannula mount for looseness or separation and if present, discontinue use of the USM and contact representative to schedule a site visit. Customers are instructed to: 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Retain a copy of letter, place with your affected system, and keep a copy of the acknowledgement form for your files. For Questions: North/South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. Europe/Middle East/Asia/Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2046-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology