Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement. Recall
Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2050-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 20, 2026
- FDA report date
- May 13, 2026
- Quantity in commerce
- 30 units
- Distribution
- See key details
- FDA cause
- Labeling mix-ups
What is being recalled?
Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
Affected model & UDI checker
Search the exact identifiers published for Z-2050-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PDSXL5 | Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement. | Check source notice | Not specified |
| Not specified | Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement. | Check source notice | 00843193113979 |
FDA code information
Model Number: PDSXL5; UDI-DI: 00843193113979; Lot Number: 2026-0026
Why was this product recalled?
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
What actions did Centinel Spine, Inc. take?
Centinel Spine notified distributors on about 03/20/2026 via telephone calls and emailed letter. Surgeons were sent letters via FedEx on 03/24/2026 and by email on 03/25/2026 & 03/26/2026. Distributors were instructed to identify any affected units on hand, provide photographs documenting such, and arrange for the return of all affected units. Additionally, distributors have been instructed to diligently check prodisc C SK inventory prior to any scheduled surgery to ensure that the recalled lot numbers are not available for implantation. This includes both consignment and loaner inventory. Acknowledgment of the letter has been requested with return envelopes provided. For surgeons, they were instructed to monitor any patient(s) that have received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures. Centinel Spine intends to remove from the market all remaining implants within the lots effected by this recall and issue a replacement product with corrected labeling for any remaining products that are subject to this recall.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2050-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology