Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2050-2026

Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement. Recall

Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2050-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 20, 2026
FDA report date
May 13, 2026
Quantity in commerce
30 units
Distribution
See key details
FDA cause
Labeling mix-ups

What is being recalled?

Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.

Affected model & UDI checker

Search the exact identifiers published for Z-2050-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PDSXL5Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.Check source noticeNot specified
Not specifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.Check source notice00843193113979
FDA code information

Model Number: PDSXL5; UDI-DI: 00843193113979; Lot Number: 2026-0026

View the original FDA recall notice

Why was this product recalled?

Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.

What actions did Centinel Spine, Inc. take?

Centinel Spine notified distributors on about 03/20/2026 via telephone calls and emailed letter. Surgeons were sent letters via FedEx on 03/24/2026 and by email on 03/25/2026 & 03/26/2026. Distributors were instructed to identify any affected units on hand, provide photographs documenting such, and arrange for the return of all affected units. Additionally, distributors have been instructed to diligently check prodisc C SK inventory prior to any scheduled surgery to ensure that the recalled lot numbers are not available for implantation. This includes both consignment and loaner inventory. Acknowledgment of the letter has been requested with return envelopes provided. For surgeons, they were instructed to monitor any patient(s) that have received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures. Centinel Spine intends to remove from the market all remaining implants within the lots effected by this recall and issue a replacement product with corrected labeling for any remaining products that are subject to this recall.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.