Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; Recall
Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2068-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 7, 2026
- FDA report date
- May 13, 2026
- Quantity in commerce
- 101 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
Affected model & UDI checker
Search the exact identifiers published for Z-2068-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880341 |
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880358 |
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880365 |
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880372 |
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880389 |
| Not specified | Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; | Check source notice | 04260495880396 |
FDA code information
Software Versions: 3.4.0 up to 3.9.5. UDI-DI: 04260495880341, 04260495880358, 04260495880365, 04260495880372, 04260495880389, 04260495880396.
Why was this product recalled?
If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.
What actions did Mint Medical GmbH take?
On or about January 7, 2026 Urgent Field Safety Notice letters were emailed to customers. Customers were instructed to read the notice and check if any affected product is available for use. If that is the case, the malfunction may have occurred and may occur in your system. Please be aware that the malfunction may occur. Your IT department must ensure that no temporary network interruptions occur in your clinical network. Also, your IT department must ensure that the mint Lesion server is not restarted while users work with mint Lesion application (i.e., the server may be restartet, but only outside of working hours). Please verify correctness of the information of prior time-points. Be attentive to potential inconsistencies in lesion attributes when opening existing cases for follow up. Cross check lesion location and categorization with prior radiological reports and image data. If you suspect that the problem may have occurred on your mint Lesion system, please contact Mint Medical support via [email protected] or any other contact details for technical analysis and investigation. Mint Medical Support will analyze if the malfunction appeared at your mint Lesion system in the past. If you believe that this failure has occurred, please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Pass this notice on to all users of mint Lesion" within your organization. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2068-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology