Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2077-2025

ColoSense Test Kit, Part No. 80-001, component of ColoSense test Recall

ColoSense Test Kit, Part No. 80-001, component of ColoSense test

FDA recall Class II

Ongoing

FDA lists this Class II recall for ColoSense Test Kit, Part No. 80-001, component of ColoSense test. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2077-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 20, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
10 units
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

ColoSense Test Kit, Part No. 80-001, component of ColoSense test

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Lot number 80-001-A2501

View the original FDA recall notice

Why was this product recalled?

Multiple plate failures documented as a result of the Low Positive Template Control being too high and out of range using the affected lot of ColoSense Test Kit.

What actions did Geneoscopy, Inc. take?

On May 20, 2025, the firm initiated a voluntary removal of the affected test kit lot. An internal recall notification was sent to the affected CLIA Laboratory Director. All remaining test kits were removed from the CLIA laboratory. The firm's investigation of the root cause for this issue is ongoing.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology