Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2081-2025

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Recall

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2081-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 14, 2025
FDA report date
Jul 16, 2025
Quantity in commerce
13240
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-2081-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
OPTI18Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial CannulaCheck source noticeNot specified
Not specifiedBrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial CannulaCheck source notice00690103180565
FDA code information

Lot Code: Model No: OPTI18; UDI: 00690103180565; All lots up to: BSLC9155 Serial Numbers: 130451 130466 130472 130593 130709 130764 130818 130821 130847 130948 130952 131038 131043 131047 131182 131193 131231 131236 131289 131311 131317 131433 131435 131513 131516 131553 131554 131606 131623 131625 131628 131682 131737 131909 131929 131966 132175 132201 132366 132395 132416 132430 132435 132505 132832 132958 132959 BSLC0110 BSLC0166 BSLC0661 BSLC0661 BSLC0826 BSLC0826 BSLC0826 BSLC0826 BSLC0855 BSLC0855 BSLC0887 BSLC0887 BSLC0979 BSLC1114 BSLC1114 BSLC1211 BSLC1308 BSLC1469 BSLC1656 BSLC1660 BSLC1815 BSLC3221 BSLC4153 BSLC4400 BSLC4461 BSLC4513 BSLC5027 BSLC5027 BSLC5320 BSLC5367 BSLC5765 BSLC6421 BSLC6467 BSLC7335 BSLC7514 BSLC7515 BSLC8175 BSLC8385 BSLC8535 BSLC8657 BSLC9100 BSLC9104 BSLC9105 BSLC9152 BSLC9155

View the original FDA recall notice

Why was this product recalled?

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

What actions did Edwards Lifesciences, LLC take?

On 05/16/2025, and 05/19/2025, the firm sent via FedEx overnight an "URGENT: PRODUCT RECALL" Letter to customers informing them that Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. The risk for the wire-reinforcement coil protruding out of the cannula body is major tissue damage and hemolysis. Customers are instructed to: " Verify their inventory on the attached customer acknowledgement form. " Share this notice with appropriate clinical staff at their site. " No patient follow-up or notification is necessary. " Return a completed Customer Acknowledgment Form to Edwards eCV Representative or via email to Edwards Customer Service at [email protected] within 15 days of receipt of this notification. " After receiving RGA number, return any impacted product to Edwards at the address below. A credit will be issued upon receipt of returned product. enableCV Distribution Center 6644 W. 2100 S. Suite D West Valley, UT 84128 For questions/assistance contact - Customer Service at 1-888-943-2783, or email: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology