Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2091-2025

TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances Recall

TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances

FDA recall Class III

Completed

FDA lists this Class III recall for TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2091-2025
Recall classification
Class III
Enforcement status
Completed
Initiation date
Jan 19, 2024
FDA report date
Jul 16, 2025
Quantity in commerce
100
Distribution
See key details
FDA cause
Software design

What is being recalled?

TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances

Affected model & UDI checker

Search the exact identifiers published for Z-2091-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
OBJ-09165TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliancesCheck source noticeNot specified
Not specifiedTrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliancesCheck source notice7290018915400
FDA code information

UDI-DI 7290018915400 Lots 23485 (exp. 08-Nov-2024) and 22904 (exp. 09-Feb-2024)

View the original FDA recall notice

Why was this product recalled?

Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.

What actions did STRATASYS LTD take?

The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology