FDA RECALL RECORD · Z-2096-2025

Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2 Recall

Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2

FDA recall Class I

Ongoing

FDA lists this Class I recall for Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2096-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jun 20, 2025
FDA report date
Jul 23, 2025
Quantity in commerce
28 units
Distribution
See key details
FDA cause
Employee error

What is being recalled?

Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2

Affected model & UDI checker

Search the exact identifiers published for Z-2096-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
35700BAX2Baxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2Check source noticeNot specified
Not specifiedBaxter SIGMA Spectrum Infusion Pump V8, Model Number 35700BAX2Check source notice00085412498683
FDA code information

UDI-DI: 00085412498683; Serial Numbers: 2006307 , 2052091 , 2093765 , 2095360 , 2104302 , 2119236 , 2141815 , 2025925 , 2067699 , 2094483 , 2095541 , 2105238 , 2119436 , 2143406 , 2034297 , 2074097 , 2094580 , 2096351 , 2105769 , 2135482 , 2143684 , 2051591 , 2090658 , 2095202 , 2096856 , 2118940 , 2137976 , 2156201

View the original FDA recall notice

Why was this product recalled?

Certain Spectrum infusion pumps may have an incorrect version of software.

What actions did Baxter Healthcare Corporation take?

An URGENT MEDICAL DEVICE RECALL notification letter dated 6/20/25 was sent to customers. Actions to be Taken by Customers 1. Immediately locate Spectrum pumps with the affected serial numbers and remove them from service. The product code and serial number can be found on the bottom of the infusion pump. 2. Check and record the software version installed on the pump. The software on a V6 pump should begin with the number 6 and the software on a V8 pump should begin with the number 8. The software version is located on the startup screen upon powering up the pump as pictured on the next page. Additionally, the software version is indicated in the pump menu which can be accessed by following the steps. For V6: From the Pump Information screen, press the sw info soft key to display the software version screen. For V8: From the Select Care Area screen, press the options menu soft key. Select User Options and press OK. Select View Information and press OK. Select Pump Information and press OK to display the Pump Info screen. 3. Contact Baxter to confirm the software version on the affected pumps and to schedule service if required. Baxter Global Technical Services can be reached at 800-843-7867 (select option 1, then option 2), Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. 4. If you received this communication directly from Baxter, acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology