SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000 Recall
SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
FDA recall Class II
Ongoing
FDA lists this Class II recall for SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2116-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 3, 2025
- FDA report date
- Jul 23, 2025
- Quantity in commerce
- 1,445,600
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
Affected model & UDI checker
Search the exact identifiers published for Z-2116-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| SCXT3-5-2000 | SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000 | Check source notice | Not specified |
| SCXT3-2000 | SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000 | Check source notice | Not specified |
FDA code information
REF/UDI-DI/Lot(Expiration): SCXT3-5-2000/10887709097107/14258739(1-Dec-26), 14258743(1-Feb-26), 14258744(1-Dec-26), 14264200(1-Dec-26), 14272241(1-Dec-26), 14272248(1-Dec-26), 14287456(1-Dec-26), 14287459(1-Dec-26), 14287480(1-Dec-26), 14287482(1-Dec-26), 14287485(1-Dec-26), 14294832(1-Dec-26), 14272254(1-Dec-26), 14310313(1-Dec-26), 14310425(1-Dec-26), 14323354(1-Dec-26), 14287475(1-Dec-26), 14294840(1-Dec-26), 14321645(1-Jan-27), 14321652(1-Jan-27), 14331169(1-Jan-27), 14332773(1-Jan-27), 14341732(1-Oct-26), 14294825(1-Jan-27), 14303209(1-Jan-27), 14310308(1-Jan-27), 14355404(1-Jan-27), 14355409(1-Jan-27) SCXT3-2000/20887709079162/14258214(1-Jan-27), 14258206(1-Jan-27), 14258208(1-Feb-27).
Why was this product recalled?
Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Icu Medical, Inc. take?
On 6/3/2025, correction notices were mailed and emailed to customers who were asked to do the following: 1) Inform all potential users of the product in your organization of this notification. 2) Discard the affected devices following your institution's process for discarding product. If discarding the device is not immediately possible at your facility, the device should be quarantined until disposal is possible. 3) Complete and return the Customer Response Form to [email protected] 5. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. For further inquiries, please contact firm at the following: Global Complaint Management, [email protected],1-(866)-216-8806; Customer Service, [email protected], 1-(800)-258-5361; Field Action Processing, [email protected], 1-(866)-328-0114
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2116-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026