Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2120-2025

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Recall

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2120-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 10, 2025
FDA report date
Jul 23, 2025
Quantity in commerce
37429 units (US)/11,703 units (OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component

Affected model & UDI checker

Search the exact identifiers published for Z-2120-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
806Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier ICheck source noticeNot specified
FDA code information

Lot Code: Product Code: 806 7324 (UDI: 10758750004317) Lots currently within expiry: V1665 - Exp.8/14/2025 B2245 - Exp 3/25/2026 Y2110 - Exp. 3/25/2026 E2828 - Exp. 2/3/2027

View the original FDA recall notice

Why was this product recalled?

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?

On June 10, 2025, "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letters were sent to distributors and customers. CUSTOMER LETTERS: REQUIRED ACTIONS " Monitor VITROS CHE Slides QC results to detect a drift in VITROS Performance Verifiers I and II below 2x Within-lab SD limits. If QC results are below 2x WLSD, perform QC again using freshly reconstituted vials of VITROS Performance Verifiers I and II. " Complete the enclosed Confirmation of Receipt form no later than June 18, 2025. " Save this notification with your User Documentation or post this notification near your laboratory's fluid reconstitution location until the issue has been resolved. " Please forward this notification if the affected products were distributed outside of your facility. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization. QuidelOrtho's investigation is ongoing, and we are working to identify root cause. We will communicate again after root cause has been determined and the issue has been resolved. Please note that until this issue is resolved, future lots of VITROS Performance Verifiers I and II may not meet the 7-day reconstituted stability for CHE. DISTRIBUTORS: REQUIRED ACTIONS " Please send the enclosed customer letter and customer Confirmation of Receipt form to all customers who have been shipped VITROS Chemistry Products Performance Verifiers AND VITROS CHE Slides from August 1, 2023, to date. " Complete the enclosed distributor Confirmation of Receipt form no later than June 18, 2025. " Please forward this notification if the affected products were distributed outside of your facility. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.