KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L Recall
KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L
FDA recall Class II
Ongoing
FDA lists this Class II recall for KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2124-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 17, 2025
- FDA report date
- Jul 23, 2025
- Quantity in commerce
- 21 total
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L
Affected model & UDI checker
Search the exact identifiers published for Z-2124-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 0805K | KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L | Check source notice | Not specified |
FDA code information
Lot #: 805-234-4, 805-234-5 UDI: (0805K) 30845357019531, (0805L) 10845357019544
Why was this product recalled?
Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.
What actions did Microbiologics Inc take?
On June 17. 2025, the firm began notifying affected customers via "Urgent Medical Device Recall" letters delivered through email. Customers were instructed to review lab procedures to understand how the information may affect usage; use or discard product depending on lab procedures. Complete and return the response form to [email protected]. If product was further distributed or transferred to other facilities or customers, they should also be notified. Customers may contact Microbiologics if a replacement kit is needed. If you have any questions or concerns, contact Recall Support team at 320.229.7080 or [email protected]. Collect calls may be made.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2124-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology