Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2131-2026

BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC Recall

BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC

FDA recall Class II

Ongoing

FDA lists this Class II recall for BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2131-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 25, 2026
FDA report date
May 13, 2026
Quantity in commerce
1266
Distribution
See key details
FDA cause
Device Design

What is being recalled?

BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC

Affected model & UDI checker

Search the exact identifiers published for Z-2131-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ORB-0320C-KCBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source noticeNot specified
ORB-0310C-KCBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source noticeNot specified
ORB-0304C-KCBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source noticeNot specified
ORB-0302C-KCBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source noticeNot specified
ORB-0305D-KCBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source noticeNot specified
Not specifiedBioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KCCheck source notice04573190050552
FDA code information

UDI-DI: 04573190050552. Lots: 25043001, 25043002, 25043003, 25051401, 25051402, 25051701, 25051702, 25060701, 25060702, 25060703, 25060704, 25083001, 25083002, 25083003, 25083004, 25083005, 25120301, 25122701, 25122702, 26020401 25F07101AA, 25F07103AA, 25F07106AA, 25H27105AA, 25H27107AA, 25H27102AA, 25H27103AA, 25H27104AA, 25K08102AA, 25K08103AA, 25K08101AA, 25J27104AA, 25J27103AA, 25J27102AA, 25J27101AA, 25H27108AA, 25H27101AA

View the original FDA recall notice

Why was this product recalled?

Resorbable bone void filler falls outside standard specifications.

What actions did Orthorebirth Co Ltd take?

On 3/25/2026, product withdrawal emails were sent to customers informing them to stop distribution and that additional information on device return would be provided. If products were distributed to distributors or hospitals, the firm requests the customer inform to stop distribution and ensure affected devices are not used. In addition, notify the firm if affected devices were used in surgery. On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1) Return affected devices to firm. 2) If devices were distributed elsewhere, recover all affected product that is in the hands of sub-distributors or end-users, and that you return all affected product to firm. 3) Complete and return the customer reply form via email to [email protected] If you have recall questions, an alternative firm contact for the recall can be found at 1-858-792-1235

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology