Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Recall
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3¿ cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan¿ helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2136-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 19, 2026
- FDA report date
- May 20, 2026
- Quantity in commerce
- 480
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3 cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-2136-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 11-1424 | Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH | Check source notice | Not specified |
| Not specified | Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH | Check source notice | 00888912017305 |
FDA code information
Model/Catalog Number: 11-1424 Lot Code: Lot: 110525 GTIN: 00888912017305
Why was this product recalled?
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Djo, LLC take?
On 03/19/2026, the firm emailed an "Urgent: Medical Device Recall" Letter to customers to inform them that, twenty-nine complaints have been received reporting a consistent failure mode described as a broken connector. The connector is the component that attaches the end of the circulating hose from the cooler to the pad which is placed on/around the affected limb. Excess water leakage from this component is a key indicator of breakage. Customers are instructed to: Enovis (DJO, LLC) requires customers to take the following actions by 03-31-2026: This notice must be forwarded to all those within their organization or to any organization where the affected product has been transferred. 1. Review stock of the affected products and lot numbers to be returned. Quantities or affected product previously shipped are listed in the table at the opening of this letter. 2. Quarantine these devices from sale. 3. Complete the attached acknowledgment and response form and return to ([email protected]). 4. To return affected product, please contact ([email protected]) or call (800) 336-6569 option 4 and customers will be provided with a shipping label to return the affected product. 5. Return all affected products using the shipping label provided. For questions - contact Kenneth Newman at [email protected] or call/text 682-299-9133.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2136-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology