FDA RECALL RECORD · Z-2140-2025

Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Recall

Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2140-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 18, 2025
FDA report date
Jul 30, 2025
Quantity in commerce
41 units
Distribution
See key details
FDA cause
Packaging change control

What is being recalled?

Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

Affected model & UDI checker

Search the exact identifiers published for Z-2140-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CS-G7Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR.Check source noticeNot specified
Not specifiedGreen Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR.Check source notice00821925038332
Not specifiedGreen Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR.Check source notice00821925001190
FDA code information

Model Number: CS-G7. UDI-DI: 00821925038332, 00821925001190. All lot numbers

View the original FDA recall notice

Why was this product recalled?

Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 06/18/2025 via letter. Consignees were instructed to review the notification and arrange for the return of all affected units on hand, ensure all affected personnel are aware, and notify customers if the affected units were further distributed. Olympus also requested that consignees complete and return the response form acknowledging receipt of the notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.